1,1-二氯乙烷CH3CHCl2氣體檢測管 235SA 1,1- 10-160ppm 乙叉二氯,1,1-Dichloroethane,使用時將管子兩端割斷,讓含有被測物質的氣體定量地通過管子,被測物質即和管中檢測劑發生定量化學反應,部分檢測劑被染色,其染色長度與被測物濃度成正比,從檢測管上已印制好的刻度即得知被測氣體的濃度。
更新時間:2024-12-18
在線留言1,1-二氯乙烷CH3CHCl2氣體檢測管1,1-Dichloroethane 235SA 1,1- 10-160ppm
Tube No.235SA
1. PERFORMANCE
1) Measuring range :10-160 ppm
Number of pump strokes 1(100mr)
2) Sampling time :2 minutes/1 pump stroke
3) Detectable limit :3 ppm
4) Shelf life :1 year(Necessary to store in a refrigerated place ; 0~10℃)
5) Operating temperature :0~40℃
6) Operating Humidity :20-90 R.H.%
7) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
8) Reading :Direct reading from the scale calibrated by 1 pump stroke
9) Colour change :White→Purple
2. RELATIVE STANDARD DEVIATION
RSD-low:15% RSD-mid.:15% RSD-high:10%
3. CHEMICAL REACTION
Chlorine is produced by decomposing with an Oxidizer.
By reacting between this Chlorine and 3, 3’-Dimethylnaphthidine,Nitroso-compound is produced.
CH3CHCI2+CrO3+H2SO4→CI2
CI2+3, 3’-Dimethylnaphthidine→Nitroso-compound
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
????
GAS ???? To pump
DETECTOR TUBE
(Olive)
PRETREAT TUBE
160
140
120
100
80
60
40
20
10
ppm
100ml
235S 1,1-DICHLOROETHANE
Substance Interference ppm Coexistence
Nitrogen oxides Similar stain is produced. Higher readings are given.
Halogens 〃〃
Halogenated hydrocarbons 〃〃
Alcohols Tafhfeec atecdc.uracy of readings is not 400 Lower readings are given.
Hexane FIG.1 〃20 〃
Toluene FIG.2 〃20 〃
0
100
200
Hexane〔ppm〕?
FIG.1 Influence of Hexane
1,1-Dichloroethane indicating value 〔ppm〕
50 100 0
100
200
Toluene〔ppm〕?
FIG.2 Influence of Toluene