丙酮氣體檢測管102SD ACETONE (-) p-Cymene CH3C6H4CH(CH3)2 20-200ppm,環境監測站近日從我公司購入一套惡臭氣體應急檢測箱,用于檢測各種場合大氣中的惡臭氣體和臭氣濃度。這一儀器將替代國標“三點式臭袋法",其檢測效果等同于“三點式臭袋法",而檢測效率將獲明顯提升。
更新時間:2024-12-18
在線留言丙酮氣體檢測管102SD ACETONE 20-200ppm
Tube 02SD
1. PERFORMANCE
1) Measuring range :125-5,000 ppm 50-2,000 ppm 20-800 ppm
Number of pump strokes 1 / 2(50mL) 1(100mL ) 2(200mL)
2) Sampling time :1.5 minutes / 1 pump stroke
3) Detectable limit :10 ppm(200mL)
4) Shelf life :2 years
5) Operating temperature :0~40℃
6) Temperature compensation:Necessary( See "TEMPERATURE CORRECTION TABLE" )
7) Reading :Direct reading from the scale calibrated by 1 pump stroke
8) Colour change :Yellow→Dark brown
2. RELATIVE STANDARD DEVIATION
RSD-low :10 % RSD-mid.:5 % RSD-high :5%
3. CHEMICAL REACTION
Chromium oxide is reduced.
CH 3COCH 3+Cr6++H 2SO 4→Cr3+
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
102SD ACETONE
Substance Interference Coexistence
Alcohols Brown stain is produced. Higher readings are given.
Other Ketones 〃〃
Esters 〃〃
Aromatic hydrocarbons 〃〃
Paraffin hydrocarbons
If the top of discoloured
layer is possible to read,
the accuracy of readings is
not affected.
〃
(more than C )
Halogenated hydrocarbons 〃〃
(NOTE)
In case of 1 / 2 or 2 pump strokes, following formula is available for the actual concentration.
1 / 2 stroke:s Actual concentration = 2.5 ×Temperature corrected value
2 stroke:s Actual concentration = 0.4 ×Temperature corrected value
TEMPERATURE CORRECTION TABLE
Scale True Concentration (ppm)
Readings 0 ℃ 10 ℃ 20 ℃ 30 ℃ 40 ℃
(ppm) (32 °F) (50 °F) (68 °F) (86 °F) (104°F)
2,000 3,500 2,800 2,000 1,600 1,300
1,500 2,750 2,000 1,500 1,250 1,000
800 1,300 1,000 800 700 600
600 1,000 800 600 550 450
400 650 500 400 350 300
200 300 200 200 200 200
100 200 100 100 100 100
:
:
50 100 50 50 50 50
1,000 1,670 1,250 1,000 850 700
GAS To pump
DEHUMIDIFIER
(White)
3
www.r-biopharm.com.cn